emperature Range: 50 to 250°C
Temperature Resolution: 0.1°C
Stability: ±0.015°C at 50°C, ±0.025°C at 150°C, ±0.05°C at 250°C
Uniformity: ±0.02°C at 50°C, ±0.04°C at 150°C, ±0.06°C at 250°C
Provide superior thermal environment for probe immersion as no air gap exist between the probe and the medium Thermal coupling is therefore much better than the alternatives described, and the stirring results in very even heat distribution throughout the medium. Methanol is used for temperature below 0°C, water from 0 to 80°C and silicon oil for upto 250°C. We have seven types of liquid baths given below.
CALsys 250 calibration bath is a highly stable standard stirred liquid bath for calibrating RTDs / Thermocouples and other temperature sensors in the laboratory. The temperature of the bath is set and controlled by a self tuned PID controller with automatic super fine adjustment. The unique features is portability with very high stability and uniformity.
The medium used in the bath is silicon oil.
Temp Range | 50 to 250°C |
Temperature Resolution | 0.1°C |
Stability | ±0.015°C at 50°C ±0.025°C at 150°C ±0.05°C at 250°C |
Uniformity | ±0.02°C at 50°C ±0.04°C at 150°C ±0.06°C at 250°C |
Controlling Sensor | Precision PRT (PT-100) |
Method of control | Digital Self Tuned PID Controller |
Time to Reach Min. Temp | 2 Hrs |
Medium of Heating | Silicon Oil |
Controlling Sensor | PT-100 |
Time to Reach Max Temp | 30 Mins |
Computer Interface | RS - 232 |
Operating Temperature | 20 to 25°C |
Power Requirement | 230 VAC, 1.0 KW |
Dimensions | 360(H) x 185(W) x 285(D) mm |
Access Opening | 90 mm Dia |
Depth | 140 mm |
Removable Insert Construction | 6 holes of 15 mm Dia |
Instrument Weight | 12 Kg (without packing) |
Volume | 900 ml |
Wetted Parts | Aluminium |